Topic 9' Image Quality in Nuclear Medicine - PowerPoint PPT Presentation

1 / 43
About This Presentation
Title:

Topic 9' Image Quality in Nuclear Medicine

Description:

... amplitude versus frequency is an audio equipment's frequency response curve. ... Good low frequency response is needed to outline the coarse details of the image ... – PowerPoint PPT presentation

Number of Views:268
Avg rating:3.0/5.0
Slides: 44
Provided by: CSU157
Category:

less

Transcript and Presenter's Notes

Title: Topic 9' Image Quality in Nuclear Medicine


1
Topic 9. Image Quality in Nuclear Medicine
  • Spatial Resolution, Contrast and Noise
  • Evaluation of Detection and Observer Performance
  • Quality Assurance of Imaging Instruments

2
Spatial Resolution
  • Spatial resolution refers to the ability of
    imaging instrument to provide the sharpness or
    detail of images.
  • Factors affecting spatial resolution include
    collimator resolution (the main factor in nuclear
    medicine). System sensitivity requires certain
    diameter of the collimator holes etc.

3
Spatial Resolution
  • Factors affecting spatial resolution also include
    intrinsic resolution (due to the statistical
    variation which is photon energy dependent).
  • Patient motion in respiratory or cardiac imaging
    causes image blurring
  • Image display or recording system can also
    contribute the degrading of spatial resolution.

4
(No Transcript)
5
Methods for Evaluating Spatial Resolution
  • Organ phantom measurement (qualitative) such as
    brain phantom.
  • Bar phantom measurement (quantitative). There are
    a number of phantoms four quadrant bar phantom,
    parallel-line phantom, orthogonal hole phantom.
  • Point or Line Spread Functions.

6
(No Transcript)
7
Bar Phantom and Line Spread Function
  • The smallest bar space distinguishable can be
    regarded as the system spatial resolution
  • The FWHM of a point or line spread function is
    also a parameter for the spatial resolution.
  • The FWHM of the spread function is about 1.4-2
    times the resolvable bar pattern

8
(No Transcript)
9
Frequency Response Curve
  • An graph of relative output amplitude versus
    frequency is an audio equipments frequency
    response curve.
  • Inexpensive audio systems generally produce the
    mid-range audio frequencies (poor response in
    low and high frequencies)

10
(No Transcript)
11
Modulation Transfer Function
  • An image system responds to the input image
    signal (spatial frequencies) is the modulation
    transfer function (contrast is the amplitude).
    MTF(?)Cout(?)/Cin(?) where Cin(Imax-Imin)/(Imax
    Imin)
  • A typical nuclear medicine image system transfers
    lower image spatial frequencies.

12
(No Transcript)
13
(No Transcript)
14
Modulation Transfer Function
  • Good low frequency response is needed to outline
    the coarse details of the image and is important
    for the presentation and detection of relatively
    large but low contrast lesion.
  • Good high frequency response is necessary to
    portray fine details and sharp edges.

15
Modulation Transfer Function
  • Systems spatial resolution can be described
    completely by the Modulation Transfer Function
    (MTF)
  • MTF is obtained by mathematical analysis of the
    line or point spread functions
  • Minimum visible bar patterns (frequency) can be
    related to the basic spatial frequency in MTF
    which drops below about 0.1.

16
Modulation Transfer Function
  • MTF curves can be obtained for different
    components of an imaging system and the system
    MTF is then given by MTF(?)MTFi(?)xMTFc(?)
  • The MTFs can be used for the comparison of
    different systems.

17
(No Transcript)
18
Contrast
  • Image contrast refers to the differences in
    density (or intensity) in parts of the image.
  • There are a number of factors that could affect
    the contrast such as the radiopharmaceuticals
    (high lesion-to-background uptake desirable)
  • Film contrast (transparency has better contrast
    than Polaroid film) enhance both desire image
    contrast and noise.

19
Contrast
  • Background count rates can reduce the image
    contrast substantially.
  • Scattered radiation and septal penetration have
    the same effect of adding background to the
    image.
  • Pulse height analyzer (narrow the window) could
    be used for the scatter rejection but there is a
    trade-off (decrease counts and increase noise)

20
Contrast
  • Decreased contrast (background, scatter or septal
    penetration) results in poorer visibility of both
    large low contrast objects as well as fine
    details (all structures) in the image.
  • Scatters add long tails to the spread function,
    suppress the low frequencies and shift the
    limiting high frequency in MTF.

21
(No Transcript)
22
(No Transcript)
23
(No Transcript)
24
(No Transcript)
25
Background Removal Example
26
Noise
  • Image noise may be either random or structured
  • Structured noise refers to non-random variation
    in counting rate superimposed on and interfering
    with the perception of the structures of interest
  • Structured noise may arise from the radionuclide
    distribution itself or caused by system
    artefacts.
  • Random noise is caused by statistical variation
    of count rate and is very important factor in
    nuclear medicine.

27
(No Transcript)
28
Random Noise
  • Random noise is related to the information
    density
  • Information density is defined as the counts per
    unit area recorded
  • Information density can be increased by increase
    count rate or imaging time
  • Information density affects the minimum
    detectable size and contrast of lesions.

29
(No Transcript)
30
Noise vs Lesion Contrast
  • Noise contrast is the percentage standard
    deviation of counts recorded in an area
  • A 3-5 times the noise contrast is required for a
    lesion to be detectable
  • Lesion contrast requirement increases as lesion
    size decrease
  • Random noise may be the detection limiting factor
    for small, low contrast lesions.

31
(No Transcript)
32
Evaluation of Detection and Observer Performance
  • Contrast-Details Studies (C-D)
  • Receiver Operating Characteristics (ROC)

33
Contrast-Detail Studies
  • A phantom with a set of objects of varying sizes
    and contrasts is used to acquire images from
    different systems. An observer is then given the
    images randomly to identify the smallest size
    visible at each level of contrast.
  • Useful for comparing the detectability of both
    large low contrast lesions as well as small
    high-contrast lesions.

34
(No Transcript)
35
(No Transcript)
36
Receiver Operating Characteristic Studies
  • A set of images is obtained with different image
    systems or techniques to be tested. A observer is
    asked to identify positive or negative with 4
    levels of confidence.
  • True positive fraction (TPF) and False positive
    fraction (FPF) are then calculated
  • The ROC curves (TPF vs FPF) are then plotted for
    each system for comparison.

37
(No Transcript)
38
Quality Control Tests
  • Operational checks and acceptance testing
  • Quality parameters uniformity, spatial
    linearity, spatial resolution, count rate
    response, system sensitivity, multi-window
    spatial registration, energy resolution etc.
  • Quality control testing methods common sources
    99mTc and 57Co. Intrinsic and extrinsic

39
(No Transcript)
40
Acceptance Tests
  • All parameters should be measured as the
    reference for the operational checks.
  • Acceptance tests are more complicated than the
    routine tests but some of the tests are used for
    the operational checks.

41
(No Transcript)
42
Quality Control Program
  • Physicists or Technologists are responsible.
    Quality control programs depend on every
    individual institution. There are general
    standard guidelines such NEMA standard.
  • Current state-of-art computers and software
    programs are available for quality control
    purpose.

43
(No Transcript)
Write a Comment
User Comments (0)
About PowerShow.com