Title: A Detailed Introduction to the Basic Operation of SerialEM
1A Detailed Introduction to the Basic Operation of
SerialEM
2Image Buffers in SerialEM
- Images are saved in buffers, referred to as A
N - Camera acquisitions and processed images go into
buffer A, and the top few images are rolled into
higher buffers - Typical situations (enforced during tilt series)
are
Normal operation
Low dose mode
(Home) (Insert) (Delete) (End)
(Home) (Insert) (Delete) (End)
A B C D E G H
New image -gt
New image -gt
A B C D E F G H
Align reference
Record reference
Read-in image -gt
Trial reference
Read-in image -gt
3Image Display Panel
- The available data range in the displayed image
is controlled by the Black and White levels
everything outside this range is truncated - To determine Black and White, it analyzes pixels
in a central area and truncates a certain
fraction - Contrast and brightness are of limited usefulness
because of this truncation - Most useful option is to set Area Fraction to
smaller value, like spot metering on smaller
central area - To see contrast at extreme end of range, enter
image minimum as Black value or image maximum as
White value - Zoom pretty much takes care of itself with
default settings
4Microscope Status Panel
- Screen current is more useful than exposure time
Float button floats a screen meter, good when
tuning scope - Dose button floats a cumulative dose meter
- The Defocus is based on a microscope readout, not
the actual focus on the specimen - On Tecnai, it matches TUI defocus reading, can be
reset in TUI or SerialEM - On JEOL, it is arbitrary and set to zero on
program startup, can be reset - IS is image shift in microns, important to know
on 300KV Tecnai and JEOL 2100 without high power
image shift
5Tilt Control Panel
- The delay time is the delay after tilting by
basic increment delay will be longer or shorter
for bigger or smaller tilts
Camera and Macro Tools
- Buttons for camera parameter setup and 5 camera
acquisition modes (good for low dose) - STOP is the one general stop button for all kinds
of operations - Preview morphs into Montage when montaging,
unless low dose is on - Macro buttons also morph into useful buttons
during tilt series
6Camera Parameter Setup
- Continuous acquisition is only mildly useful (no
access to fast camera modes) - Processing options useful for testing and
troubleshooting otherwise always use gain
normalization - Binning is used for speed. Tracking correlations
are always binned down to 512 so there is no
point taking Trials bigger than 512. - A full-width image gives fastest readout per
pixel hence the wide options for fast subarea
pictures - A separate dark reference is kept for every
binning and exposure time. - Force new dark reference next time is the way
to remove a dark reference - Averaging dark references is helpful for a
high-noise camera (not typical)
7Pre-exposure
- Pre-exposure (drift settling) requires two
shutters - The Gatan shuttering mode that allows
pre-exposure is inflexible and exposes specimen
during readout so SerialEM uses beam blanking
to provide selected pre-exposure while blanking
during readout - Dual shuttering minimum exposure mode
provides for this and just does beam blanking
when drift settling is 0 so it can always be
used - But the camera timing must be calibrated so the
beam blanking does not occur during exposure time - Tietz cameras require Shutterbox II for
pre-exposure to work
8Image Alignment Focus
- Autoalign always changes microscope image shift
Clear alignment undoes that - Reset Image Shift moves the stage to keep the
same specimen point centered - Moving stage for mouse shifts avoids having to
reset image shift - Happens automatically if option selected and
movement is above either absolute (micron) or
relative (fraction of field) threshold - You can do it intentionally with Shift key
- Trimming dark borders keeps autoalign from being
thrown off by beam edge or grid bar - Used automatically in low dose and during tasks
done at lower than working mag - Will not detect less extreme contrast differences
that can throw off autoalign alternative is to
use smaller area (higher trim fraction) - Center image shift on tilt axis will give fewer
movements during tilt series but requires an
offset calibration
9Autofocusing with Beam Tilt
- Autofocus is done by taking two images with the
beam tilted in opposite directions and measuring
the shift between images - It relies on the microscope being aligned such
that the image of a specimen in the focal plane
of the objective lens does not move when the beam
is tilted.
A
Specimen in focal plane
Specimen out of focus
B
C
Image plane
Image plane
A appears here with both directions of tilt
B or C appears here depending on direction of
tilt
10Focus Parameters and Operations
- The target is the defocus it sets to after
measuring - The beam tilt needs to be big enough to get good
image displacements, but is limited by objective
aperture - Autofocus Offset lets you measure defocus at
focus different from target - Can protect against correlation failures by
focusing at zero - Can be used to focus at a defocus with more
contrast - Taking three pictures lets it subtract off drift
Drift protection - Report shift and drift takes the three pictures
and just tells you the beam-induced shift and the
drift - Report on existing reports the shift and drift
from pictures just acquired - Autofocus measures and changes to target
Measure defocus just measures without changing - Autofocus sometimes works poorly at high tilt -
Check autofocus assesses what fraction of a
known focus change can be detected - Move Focus Center lets you click on spot at
desired focus and change focus so center of field
is at that focus
11File Operations
- You can open new image files or existing files
(up to 8 now) - Opening a new file brings up File Properties
dialog, where you set the data type and items to
save in header - Anything selected for saving can be extracted
with extracttilts (options -st, -ma, or -in for
stage position, mag, or intensity) - After opening an old file, the Set options can
be used to change the properties that would be
set in the File Properties dialog - Save to Other and Read from Other can be used
to save one image or read from a file without
leaving it open - You can save an image to a specific Z value with
Overwrite - Saving a log is useful. Read Append lets you
accumulate a log across several runs of the
program
12Camera Commands
- The most commonly used item in the camera menu is
Prepare Gain Reference - Gain reference can be done at typical exposure
for non-low dose work - Having lots of counts averaged (e.g., 10 x 1/3
of dynamic range) is more important than matching
the counts for actual images - Binned gain reference works for normalizing
images with even binning useful for slow 4K
camera that is being used as 2K camera - If camera counts per electron is calibrated, this
can do automatic dose calibration - It can be useful for troubleshooting to view gain
reference or dark reference of image just taken - Post-actions controls whether image shift, mag
change, or stage move are taken during camera
readout - Requires correct camera timing parameters and
with that, it is surprisingly robust - But it can be turned off to troubleshoot problems
13Montaging The Camera Is No Limit
- To start montaging, define the number of frames
in X and Y in Montage Setup dialog - Parameters start at current state, based on
current Record parameters - The default overlap (10) is fine for image-shift
based montage - Once data has been taken, or if you open an
existing montage, most parameters in the dialog
are locked - After taking a montage, you will see overview
in buffer B, a subsampled version of the whole
area - A (composed) image of the center is left in A for
alignment - Either image can be mouse-shifted
- Prescan takes a fast highly binned montage of
the same area without storing it - Options are mostly important when using montages
as navigation maps, except - Change focus with height will take each picture
at same focus, but this may make it harder to
correct for distortions due to magnification
gradients
14Image Processing
- In general, processing works on the active image,
the processed image is put in A, and the top
images are rolled - Binned FFT bins the image to 512 pixels
- It is faster, less noisy but only shows central
part of frequency space - Probably most useful for assessing drift in image
- Live FFT will do FFT immediately in Continuous
acquisition mode (version 2.6.4?) - X-ray items are to help set up removal of X-rays
in dark references and images - X-ray removal from images should only be done for
low-dose images from low gain cameras where
X-rays screw up tracking - Show cross-corr will do an autoalignment with
current A and align buffer images, then show
correlation
15Tasks
- Tasks are (mostly) operations required for doing
tilt series, involving a sequence of stage
movements and tracking images - Eucentricity has two different routines
- Rough eucentricity will work from 100 µm away or
more in Z - Refine eucentricity does a 24 tilt series to
measure Z height with lateral offset of tilt axis
taken into account and estimated also - Works up to 10 µm away or more
- Refine Realign runs it in a way that restores
the image position at the end - Can use result to Set Tilt Axis Offset
Optic axis
Tilt axis
24
?Z
Axis offset
Specimen feature
0
-24
16Tasks 2
- Walk Up goes from one tilt to another while
keeping an image feature centered - Anchor is an image at intermediate tilt that
can be a reference when coming back down in a
tilt series - Reset IS Realign and Reverse Tilt keep an
image feature centered when resetting image shift
or reversing direction of tilt - Set Intensity lets you change intensity based
on the counts in the current image - Enter a factor to change intensity by (e.g.,
0.75, 1.5) - Or enter desired counts in a Record image (e.g.,
4000) - Move Beam - if you click at the center of the
beam, this will move the beam to the center of
the field
17The Problem of Tilt Series Acquisition
- The goal in acquiring a tilt series is to have
every picture centered on the same specimen
feature, and at the same focus - The basic sequence of operations required is
thus - Tilt to a new tilt angle
- Correct change in specimen position
- Correct change in specimen focus
- Acquire final (Record) image
- For full automation, the desired features must
stay centered during initiation and after
interruptions of the series the tasks are
components for ensuring this
18Approaches to Automated Tomography
- Traditional approach track and focus at every
tilt - Used in original Agard/Koster software and in
Tietz software - May have to track at low mag time-consuming
- Precalibration take a coarse tilt series to
determine X, Y, and focus, then take full tilt
series without tracking or focusing - Used in new Koster software for Tecnai and in
FEI software - Can be fast but is susceptible to drift and
calibrations being off hard to accommodate user
intervention - Z-prediction calibrate tilt axis location and
assume specimen moves in a circle around it (new
Agard method) - - Very fast, but many scopes do not meet the
assumptions - Robust prediction predict X/Y/Z position on
next tilt from changes in position on previous
tilts - Must adapt to non-ideal conditions and user
interventions - Goal is to achieve the reliability of
traditional approach and speed of precalibration
19The Robust Prediction Method in SerialEM
Tilt 2
Y
Actual position
Predicted position
Make and use prediction but track
Tilt 3
Y
Make and use prediction but track
Tilt 4
Y
Rely on prediction do not track
Tilt 5
Y
20Tilt Series Preliminaries
- Adjusting beam and exposure parameters
- Tradeoff between specimen damage and
signal-to-noise ratio in images and
reconstruction - For plastic section work, additional constraints
of dynamic range of camera and difficulty of
getting drift-free high-exposure image - Whether to let program start at zero or go to
high tilt manually - Manually means you refine eucentricity and run
walk-up - You would do this if
- You are not sure you can get to the highest tilt
angle - You may need to adjust drift settling at high
tilt - You are unsure of what counts you can reach at
high tilt - Pre-expose plastic sections to minimize shrinkage
and warping during series - 2000 electrons/Ã…2
21Tilt Series Setup the Mother of All Dialogs
- Tilt and focus parameters are the same as the
ones exposed elsewhere there is just one
underlying parameter - Some choices will be set/enabled depending on
whether you are at zero or already at high tilt - Low mag tracking needed only if field of view lt
1 µm the low mag only needs to be low enough to
give gt 2 µm field (all assuming well-behaved
stage)
22Tilt Series Setup 2 - Intensity Control
- Intensity control is important because otherwise
higher tilt images will have lower SNR - Intensity control is complicated because it
covers several situations - You might want constant intensity if doing cosine
tilting - 1/cosine or 1/cosine to a power is preferred for
cryo - Use a target of counts for stained material
1/cosine can give big variations in counts - The problem with constant counts is that it may
require too bright a beam at high tilt. Two
solutions - Taper counts down above an angle
- OK once you are familiar with usable values
- Use with Do not increase intensity above value
for first saved image to protect against
overconstricted beam at far end of series - Set up intensity before starting and use Keep
intensity below current value - Preferred method if you are starting at high tilt
and setting up beam there - Can be used even if starting at zero tilt as long
as images arent too bright there
23Tilt Series Setup 3 Autofocus
- The autofocus interval sets how often it will
focus anyway when predictions are reliable 5-6
is good value - Can focus every time above an angle not really
needed at end of series (should have an option to
do this only at start) - Checking autofocus is designed to prevent runaway
focus changes if autofocus doesnt work well
enough at high tilt - You can follow the advice, i.e. adjust focus
manually, and set defocus target to measured
defocus value so it will stop trying to change
focus - Using an autofocus offset that is the negative of
the target defocus might also work - The (rare) problem usually goes away by 50
24Tilt Series Setup 4 Initial Actions
- It is good practice to have a centered image
before you go into the dialog, then select the
option to align to it - You should refine eucentricity before a series,
especially after pre-irradiating the specimen - Rationale for anchor is that tracking is more
likely to drift off at high tilt and is likely to
be accurate below 45-50 - Aligning to an image taken during walkup at such
an angle will result in desired area being
centered at low tilt - There is no data loss from the misalignments at
higher tilt, since the drift is lateral to the
tilt axis and lost image area is not needed for
backprojection - You can have it take the anchor when it walks up,
or you can to Walkup Anchor before starting
series and then select the anchor there
25Tilt Series Setup 5 Tracking Control
- The defaults are good for normal work (defaults?
What defaults? It needs a Reset Defaults
button.) - Repeat Record if percentage of field lost can
be 5 for typical work - Make it smaller, down to 2.5, if area of
interest goes very close to borders at 0 - Make it bigger or turn it off for very high mag
or low dose - Maybe make it bigger for very large montages
- Get tracking image when error in X/Y prediction
gt is typically 1.5, could be bigger for high
mag or low dose work to reduce tracking - Stop if Autoalign shifts more than is a
protection against wild alignments - Get new track reference if Record alignment
differs by gt is relevant if low mag tracking or
low dose is in effect could be relaxed (made
bigger) for very high mag
26Resuming and Backing Up a Tilt Series
- A series can be stopped with End, at the end of
a cycle, or with STOP, right away. - If you resume from the Setup dialog it goes on to
the next step - The Resume Dialog gives you options for resuming
- The next action is listed
- You can repeat steps on the current cycle or
sometimes skip steps that you have taken care of
while stopped - You can shift the image in A into a new alignment
and check Use image in Buffer A as reference for
alignments - If you take a new Record it automatically
overwrites the previous one - The Backup dialog lets you back up the tilt angle
and overwrite multiple images - When you back up, it tilts the stage back, and
loads the Record from that tilt into A and the
image from the previous tilt into D, recreating
the situation at that tilt - You can open the dialog repeatedly and go forward
or backward to the desired point
27Verbose Log at Start of Series
28Verbose Log with Good Predictions