PROGRAMMING - PowerPoint PPT Presentation

1 / 59
About This Presentation
Title:

PROGRAMMING

Description:

Alter apical gains if sound different to SPRint using the gain shaper. CIS ... Electrode 22 Apical Lowest pitch percept. Frequency-to-channel allocation depends on: ... – PowerPoint PPT presentation

Number of Views:39
Avg rating:3.0/5.0
Slides: 60
Provided by: cochle
Category:

less

Transcript and Presenter's Notes

Title: PROGRAMMING


1
PROGRAMMING
2
MAPPING
  • Individual patients have different perceptions
    resulting from electrical stimulation to the
    cochlea
  • ie the strength of the stimulation required to
    elicit auditory perceptions of the same loudness
    is different from patient to patient, and from
    channel to channel for the same patient
  • Patients differ in their ability to perceive
    pitch changes resulting from stimulation of
    different channels

3
MAPPING
  • The CI can be programmed to fit the
    psychophysical requirements of individual implant
    recipients
  • The combination of all these individualised
    parameters is called a MAP
  • MAPping is a PROCESS by which the optimal
    programming parameters are determined for each
    individual patient

4
Speech Processing Strategies
5
SPEAK
  • Continuously analyzes the energy in the incoming
    acoustic signal and determines the spectral
    maxima.
  • Spectral maxima frequency regions in the
    waveform that contain the greatest amounts of
    energy.
  • Selection of only the highest amplitude spectral
    maxima increases the probability that speech
    rather than noise (static) is presented.

6
SPEAK
Analysis bandwidth 116 to 7787 Hz Select the
spectral maxima Sequentially stimulate the
channels representing those frequency bands
7
SPEAK
  • Amount of current applied to each channel is
    within the dynamic range (the difference between
    C-level and T-level).
  • Acoustic input varies continuously over time, so
    the spectral maxima will change and the channels
    selected for stimulation will vary across the
    array.
  • Average no. of maxima 6 8 (max 10)
  • Actual no. depends on signal spectral
    composition, signal intensity and the MAP.

8
SPEAK
  • The channels selected for stimulation represent
    the largest spectral peaks.
  • They are stimulated in tonotopic order from high
    to low frequency.
  • Stimulation rate 250 pps (overall rate 2500
    pps)
  • For broadband signals more maxima, slower
    stimulation rate
  • For sounds with limited spectral content less
    maxima, higher stimulation rate

9
SPEAK
10
SPEAK
  • SPEAK in SPrint
  • Stimulation rate
  • 250 Hz per channel (fixed)
  • Not affected by high T/C levels
  • Not affected by wide P.W. (lt 200 usec)
  • SPEAK in ESPrit
  • Selects 8 maxima (default)
  • Max. PW 37 usec
  • Requires at least 12 channels
  • Alter apical gains if sound different to SPRint
    using the gain shaper

11
CIS
  • Continuous interleaved sampling
  • Uses smaller no. of channels at a higher
    stimulation rate per channel to represent the
    temporal changes.
  • Channel selection is fixed.
  • You may choose to use 4, 6, 8 or 12 channels.
  • Due to the power demands of this strategy, it was
    used only with the SPRint speech processor. It
    is currently available in ESPrit using lower
    stimulation rate.
  • Monopolar modes only

12
CIS
  • You choose the no. of electrodes being used.
  • A bandwidth is assigned automatically to each
    one, following the tonotopic order of the
    cochlea.
  • During stimulation, the output amplitudes from
    each band are measured and represented as
    variations in the amount of current sent to the
    channels.
  • The current delivered is directly proportional to
    the energy in the band and is within the dynamic
    range.
  • Permissible stimulation rates per channel
  • 900, 1200, 1800 and 2400 Hz
  • (max overall rate lt 14400 Hz)

13
CIS
14
CIS
15
CIS Secondary Parameters
  • Jitter
  • Recommended 10-20 jitter.
  • May reduce pinging or bursty in onset at
    higher rates and reduce buzz percepts at lower
    rates.
  • Psychophysics
  • Must be re-measured for different stimulation
    rates.
  • Increasing stimulation rate will reduce T-level,
    but C-levels less affected.
  • Increasing number of channels may require a
    global reduction in C-levels.

16
ACE
  • Advanced Combination Encoder
  • ACE SPEAK CIS
  • Encodes important spectral and temporal speech
    cues
  • Flexibility to meet individual needs
  • No. of maxima up to 20
  • Stimulation rate per channel 250, 500, 720,
    900, 1200, 1800 or 2400 (overall up to 14400 Hz)
  • 6 maxima stimulation rate per channel 2.4 kHz
  • 20 maxima stimulation rate per channel 720 Hz
  • Monopolar modes only

17
ACE
18
COMPARISON OF STRATEGIES
19
Stimulation Modes
20
STIMULATION MODES
Electrical stimulation produces current flow
between an active and an indifferent
electrode Stimulation Mode describes the
location of the indifferent electrode relative to
the active electrode The physical separation of
the active and reference electrodes determines
the spread of electrical current, hence
controlling the area over which nerve endings are
stimulated.
21
STIMULATION MODES
22
Monopolar (MP) Stimulation
  • Active electrode is intracochlear but the
    indifferent electrode(s) is extracochlear.
  • MP1 current flows between designated active
    intracochlear electrode and the ball electrode
    (R1)
  • MP2 current flows between the active
    intracochlear electrode and the plate electrode
    (R2)
  • MP12 current flows between the active
    intracochlear electrode and both extracochlear
    electrodes (R1 and R2)

23
Monopolar (MP) Stimulation
  • Why MP modes?
  • Wide spread current flow
  • Much lower current levels for T and C-levels.
  • Lower energy consumption to produce auditory
    sensations
  • Longer battery life
  • Auditory nerve can be stimulated at a higher
    rates (good for both CIS and ACE)

24
Bipolar (BP) Stimulation
  • Current flows between two intracochlear
    electrodes
  • Active indifference pair
  • Examples
  • Active Indifferent Total
    no. of channels
  • BP 13 14 21
  • BP1 13 15 20
  • BP2 13 16 19

25
Bipolar (BP) Stimulation
  • The physical separation of these electrodes
    determines the spread of current and the area
    over which the spiral ganglion cells are
    stimulated.
  • If T and C-levels are too high, a wider bipolar
    mode should be selected.
  • As the bipolar configuration becomes wider, the
    number of available pairs or channels for
    stimulation decreases.

26
Common Ground (CG) Stimulation
  • Current flows between the designated active
    electrode and all other electrodes on the array
    inside the cochlea.
  • Allows all 22 electrodes to be used as active.
  • T and C-levels are typically lower than in BP
    mode.
  • It is used diagnostically to detect electrode
    anomalies.

27
Setting Stimulation Mode
  • Try MP12 mode first.
  • It provides the greatest flexibility.
  • BP modes can use only the SPEAK strategy.
  • CG modes can use only the SPEAK strategy.
  • CG should not be used in cases of partial
    insertions of electrode array because there may
    be a higher probability of non-auditory
    sensations.
  • Whenever you choose a different mode, T and
    C-levels must be re-measured in the new mode
    because the amount of current necessary to reach
    T and C-levels will be different in each
    stimulation mode.

28
Frequency-to-Channel Allocation
29
Frequency-to-Channel Allocation
  • Electrodes are numbered sequentially.
  • Tonotopic organization of the cochlea
  • Electrode 1 Basal Highest pitch percept
  • Electrode 22 Apical Lowest pitch percept
  • Frequency-to-channel allocation depends on
  • Speech processing strategy
  • No. of channels available for stimulation in the
    MAP

30
SPEAK
  • Maximum no. of channels 20
  • 16 frequency-to-channel allocation tables
  • Table 1-6 divide overall bandwidth into 20
    channels
  • Table 7-14 divide overall bandwidth into fewer
    channels when less than 18 active channels
    are available
  • Default Table 6 (widest frequency range)
  • 116 Hz to 7871 Hz
  • There is less frequency selectivity in MAPs with
    limited numbers of channels.

31
CIS and ACE
  • Fixed default frequency range 187 7937 Hz
  • 27 frequency tables
  • Automatically allocate a frequency table
    according to no. of channels (412) available.
  • If necessary, the high frequency cut off point
    can be reduced by lowering the frequency table
    number.

32
Electrical Stimulation Levels
33
Electrical Stimulation
  • Charge-balanced biphasic current pulses

Pulse Width
Stimulus Amplitude
0 mA
Increasing the amplitude of the pulse and/or
widening the pulse width results in more
electrical charge being delivered
34
Current Level and Pulse Width
  • Current Level is expressed in arbitrary units.
  • Current Level 1 (10 mA) - 255 (1750 mA)
  • Default pulse width (PW) 25 ms/phase
  • Narrow pulses take less time to deliver
  • Therefore, higher rates.
  • Pulse width 25 400 ms/phase
  • Maximum pulse width depends on the coding
    strategy, the stimulation mode, the stimulation
    rate and the number of maxima or channels
    selected.
  • Dont widen the pulse width unless the user does
    not receive sufficient loudness when using a
    narrow pulse at maximum current amplitude.

35
Stimulus Amplitude-to-Current Level MAPping
36
Automatic Gain Control
  • To control the gain so peaks in the acoustic
    signals are limited to values that translate into
    stimulation at C-level.
  • It rapidly decreases the gain so that the peak
    value of the signal is set to a fixed value.
  • AGC slowly increases the gain after the
    high-level peak has passed.

37
Microphone Sensitivity Control
  • It determines the minimum input signal level
    required for stimulation.
  • Higher sensitivity settings lower SPL is
    required to cause stimulation.
  • Lower sensitivity settings louder sound is
    required to cause stimulation.
  • For normal listening situations,
  • Sensitivity 8 on Sprint
  • Sensitivity 3 on ESPrit

38
Microphone Sensitivity Control
  • If microphone sensitivity is always set above or
    below the optimal settings
  • Overall C-levels in MAP are not set appropriately
  • If sensitivity is set too low
  • Gain applied to incoming signal is reduced
  • Soft sounds will not be processed and cannot be
    heard.
  • If sensitivity is set too high
  • Gain applied to low-level inputs is increased
  • Low level noise will be heard and signal-to-noise
    level will decrease.

39
Base Level
  • Default 4
  • Range 4 15 in SPEAK
  • 0 15 in ACE and CIS
  • Increasing the Base Level
  • Increases the level of sound required to initiate
    electrical stimulation, and then reduces the
    background noise.
  • Reduces the dynamic range
  • Dont raise the Base Level unless it is
    necessary because it reduces the operating range
    of the processor.

40
Q-value
  • Controls the steepness of the amplitude growth
  • Determines the percentage of the electrical
    dynamic range that is allocated to the top 10 dB
    of the speech processors channel amplitude
    range.
  • The lower the Q-value, the steeper the amplitude
    growth function at the lower end.
  • Default 20
  • 80 of electrical dynamic range is MAPped into
    the top 10 dB of the processors range.
  • The lower 20 of the individuals dynamic range
    is MAPped into the lower 20 dB of the processors
    range.

41
Q-value
  • Lowering Q-value
  • Making soft sounds, including background noise,
    seem louder.
  • Raising Q-value
  • Soft voices may not be heard. May reduce a
    persons overall speech understanding ability.
  • 20 50 for SPEAK
  • 10 50 for ACE and CIS

42
Autosensitivity Control (ASC)
  • Reduce background noise.
  • Looks at the troughs (noise floor) in the
    spectral enveolope.
  • Gain decreases slowly

43
Channel Gains
  • To alter the perceived sound quality of the MAP
  • Default setting 0 dB across all channels
  • Gain is applied to each channel output and
    affects the stimulation levels sent to the
    implant.

44
How do we modify the loudness percept?
45
How to modify the loudness percept?
  • Use the T/C-level modifiers to increase/decrease
    C-levels for all channels by a fixed percentage
    of the dynamic range during live-voice testing of
    the MAP.
  • Use the T/C-level modifiers to increase/decrease
    T-levels for all channels by a fixed percentage
    of the dynamic range, making the low-level
    signals louder or softer. (Note You lower
    T-levels only when you think T-levels has
    over-estimated.)
  • Lowing Q-value results in a louder perception of
    low-level signals, including background noise.

46
How to modify the loudness percept?
  • Raising the Base Level increases the minimum
    acoustic signal that results in stimulation.
  • Use the volume control on the speech processor.
  • 0 9 on SPRint
  • 0 5 on ESPrit
  • The volume control can only reduce the C-levels
    (total reduction is 25)

47
Magnet
48
Magnet
  • The magnet strength can be adjusted by turning
    the magnet insert in the external coil.
  • Magnet strength should be assessed and if
    necessary adjusted at first fitting.
  • This is critical especially for children
  • Magnet inserts strength ½, 1, 2, 3, 4 are
    available
  • Magnet strength required will depend on thickness
    of skin flap, hair, swelling, fatty tissue etc

49
Magnet
  • Magnetic pull of magnet in headset should be as
    loose as possible but still firm enough to stay
    in place with movement of the head ie not to fall
    off head

50
Magnet
  • Skin Flap Thickness
  • Receiver-stimulator will operate for separation
    of receiving and transmitting coils up to 6 to
    9mm
  • If flap is thicker than 9mm
  • the strongest magnet may not be strong enough
  • the internal and external coil coupling will not
    be as effective as it should be
  • the signal can be degraded

51
Telemetry
52
Objective Measurements
  • electrode impedance
  • compliance voltage
  • neural response telemetry (NRT)

53
Telemetry - purpose/benefits
  • Impedance Telemetry
  • Compliance Telemetry
  • NRT Telemetry
  • Measurement of electrode impedance and for
    integrity testing (Implant Test)
  • Detection of out-of-compliance conditions which
    result in abnormal loudness growth
  • Measurement of evoked auditory (neural) response

54
Electrode Impedance
  • electrode impedance is a measure of the
    opposition to electrical current flow through the
    lead wires, electrodes and tissue
  • electrode impedance calculated by dividing the
    voltage at the electrode by the current flow
    through the electrode.
  • impedance voltage/current

55
Electrical Impedance
  • can be used to detect conditions under which an
    electrode should not be used
  • HIGH IMPEDANCE either the electrode lead wire
    is broken and open circuited, or the electrode is
    not in contact with body fluids
  • SHORT CIRCUIT low impedance (in common ground
    mode) indicates that the electrode, or its lead
    wire, is in contact with another electrode or
    lead wire

56
Compliance Telemetry
  • compliance is the ability of the implant to
    deliver sufficient voltage to generate the
    desired current level
  • out-of-compliance refers to the condition where
    the maximum voltage available from the implant is
    not sufficient to generate the desired current
    level

57
Compliance Telemetry
  • if implant is out-of-compliance
  • the actual current delivered is less than the
    current level requested
  • the patient does not perceive loudness growth
    when current level is increased
  • the MAP will not provide appropriate loudness
    information

58
Compliance Telemetry
  • causes of out-of-compliance
  • high electrode impedance
  • high current level
  • WinDPS will provide warning message if implant is
    out-of-compliance during T/C-level measurements
  • resolving out-of-compliance problems
  • goal is to reduce T/C levels
  • use wide stimulation mode eg monopolar
  • increase pulse width

59
Neural Reverse Telemetry NRT
  • The stimulation of the implant causes nerve
    fibers in the spiral ganglion to fire, which
    generates an action potential.This is the NEURAL
    RESPONSE.
  • The implant records the action potential by
    amplifying signals from intracochlear electrodes.
    The action potential is encoded and transmitted
    back to the speech processor by radio frequency.
    This is the TELEMETRY.
Write a Comment
User Comments (0)
About PowerShow.com