Title: Chem. 230
1Chem. 230 10/07 Lecture
2Announcements I
- Second Homework Set Due
- Exam 2
- Next Week
- You can bring a 3 x 5 note card with notes
(front and back) to the exam - I will provide constants but no equations
- Topics Covered
- Simple Separations vs. Chromatography
- Chromatographic Theory (Basic definitions of
parameters, meaning of parameters, how to read
chromatograms, rate theory) - Intermolecular Forces Their Effects
- Optimization
3Announcements II
- Should Sign Up for Presentation Topic Today
- Todays Topics
- Optimization (last topic on Exam I)
- Gas Chromatography
- Comparison of methods
- Historical Development
- Column types
- Analytes and Samples
- Instrumentation (mobile and stationary phases,
flow control, injection?)
4Chromatographic TheoryOptimization - Overview
- How does method development work?
- Goal of method development is to select and
improve a chromatographic method to meet the
purposes of the application - Specific samples and analytes/solutes will
dictate many of the requirements (e.g. how many
solutes are being separated and in what
concentration? what is the purpose of the
separation?, what other compounds will be
present?) - Coarse method selection (e.g. GC vs HPLC and
selection of column type and detectors) is often
based on past work or can be based on initial
assessment showing problems (e.g. 20 compounds
all with k between 0.2 and 2.0 with no easy way
to increase k) - Optimization then involves making equipment work
as well as possible (or limiting equipment
changes)
5Chromatographic TheoryOptimization What are we
optimizing?
- Ideally, we want sufficient resolution (Rs of 1.5
or greater for analyte/solute of interest peaks) - We also want the separation performed in a
minimum amount of time - Other parameters may also be of importance
- sufficient quantity if performing prep scale
separation - sufficient sensitivity for detection (covered
more with instrumentation and quantitation) - ability to identify unknowns (e.g. with MS
detection)
6Chromatographic TheoryOptimization Some trade
offs
- Flow rate at minimum H vs. higher flow rates
(covered with van Deemter Equation) low flow
rate not always desired because of time required
and sometimes smaller S/N - Maximum flow rate often based on
column/instrument damage this can set flow rate - Trade-offs in reducing H
- In packed columns, going to small particle sizes
results in greater back-pressure (harder to keep
high flow) - In GC, small column and film diameters means less
capacity and greater likelihood of column
overloading - Trade-offs in lengthening column (N L/H)
- Longer times due to more column (can be
considerably longer for HPLC due to pressure
limits)
7Chromatographic TheoryOptimization Improved
Resolution Through Increased Column Length
- Example
- Compounds X and Y are separated on a 100 mm
column. tM 2 min, tX 8 min, tY 9 min, wX
1 min, wY 1.13 min, so RS 0.94. Also, N
1024 and H 100 mm/1024 0.097 mm - Lets increase L to 200 mm. Now, all times are
doubled - tM 4 min, tX 16 min, tY 18 min. So DtR
(or d) now 2 min. Before considering widths,
we must realize that N L/H (where H is a
constant for given packing material). - N200 mm 2N100 mm. Now, N 16(tR/w)2 so w
(16tR2/N)0.5 - w200 mm/w100 mm (tR200 mm/tR100 mm)(N100
mm/N200 mm)0.5 - w200 mm/w100 mm (2)(0.5)0.5 21-0.5 (2)0.5
- w200 mm 1.41w100 mm
- RS d/ave(w) 2/1.5 1.33
- Or RS 200/RS 100 d/wave (d200/d100)(w100/w200
) (L200/L100)(L100/L200)0.5 - So RS is proportional to (L)0.5
8Chromatographic TheoryOptimization Resolution
Equation
- Increasing column length is not usually the most
desired way to improve resolution (because
required time increases and signal to noise ratio
decreases) - Alternatively, k values can be increased (use
lower T in GC or weaker solvents in HPLC) or a
values can be increased (use different solvents
in HPLC or column with better selectivity) but
effect on RS is more complicated
Note above equation is best used when deciding
how to improve RS, not for calculating RS from
chromatograms
9Chromatographic TheoryOptimization Resolution
Equation
- Dont use above equation for calculating Rs
- How to improve resolution
- Increase N (increase column length, use more
efficient column) - Increase a (use more selective column or mobile
phase) - Increase k values (increase retention)
- Which way works best?
- Increase in k is easiest (but best if k is
initially small) - Increase in a is best, but often hardest
- Often, changes in k lead to small, but
unpredictable, changes in a also (for problems in
this class we will assume no change in a with
change in T or solvent composition)
10Chromatographic TheoryGraphical Representation
Smaller H (narrower peaks)
Initial Separation
Larger k - separation increases more than width
Increased alpha (more retention of 2nd compound)
11Chromatographic TheoryOptimization Back to 1st
Example
- Compounds X and Y are separated on a 100 mm
column. tM 2 min, tX 8 min, tY 9 min, wX
1 min, wY 1.13 min, so RS 0.94. Also, N
1024, kY 4.5 and a 1.13. - What change is needed in N, k, and a to get RS
1.5?
12Chromatographic TheoryOptimization 2nd Example
- tM 1 min, tX 2 min, wX 0.1 min, tY 2.1
min, wY 0.105 so RS 0.98, a 1.1, kY 1.1 - With small initial k values, increasing k helps
more - After k gt 5, only minor increases in resolution
possible
Maximum RS
Baseline Resolved
Start Point
13Chromatographic TheoryOptimization Changes in
a - I
- In GC analysis on a DB-1 (non-polar) GC column,
the compounds acetone (KOW 0.58, bp 56C)
elutes at 7.82 min while diethyl ether (KOW
7.76, bp 34.6C) elutes at 7.97 min. Peak
widths are around 0.2 min. If the unretained
time is 1.00 min., this is a difficult separation
with this column. - Occasionally, changing T to change k will also
increase a (more on this on next slide) - Suggest a column switch (aimed at increasing a to
improve the separation).
14Chromatographic TheoryOptimization Changes in
a - II
- Changes in a with T
- Example alkanes and toluene
- In Plot, most alkanes show similar temp.
retention behavior (similar slopes no overlap) - If two alkanes overlap (e.g. two branched
alkanes), there is not much chance in increasing
a (since both have same Dret/DT) - If a separation of octane and toluene had been
performed at 150, coeluting peaks would be
observed - Decreasing T would lead to improvement because
different slopes lead to a change in a
note if chromatogram started at 200C, one would
be disappointed by initial change
15Chromatographic TheoryOptimization Changes in
a - III
- In HPLC, it is possible to change the mobile
phase to affect solute solvent interactions and
retention. - For example, if molecules A and B are separated
by normal phase HPLC using 15 2-propanol/85
hexane and are found to co-elute, solvent changes
may resolve. - One might expect that changing solvent to 25
toluene 75 hexane will increase affinity of
compound B for mobile phase relative to compound
A (due to compound B being aromatic) leading to
increase retention of B
compound A
compound B
16Chromatographic TheoryOptimization Changes in
a - IV
- The two compounds below are found to give
retention times of 8.91 and 9.02 min. (aniline
and benzaldehyde, respectively) when separated
using HPLC on a C18 column using 60 methanol/40
water vs. an unretained time of 1.62 min. - There is an easy way to increase a for this
separation. How can the mobile phase be changed
to increase a?
NH2
O
17Chromatographic TheoryOptimization Some
Questions
- Indicate how the chromatograms could be improved?
18Chromatographic TheoryReview Questions
- What is the most common way to increase retention
of analytes in gas chromatography? - a) decrease flow rate
- b) decrease temperature
- c) increase flow rate
- d) use carrier gas with larger molecular weight
- Increasing the flow rate in chromatography will
increase which term in the van Deemter equation.
(Give name or term). - What type of intermolecular force is typically
the most important for analyte stationary phase
in reversed phase HPLC? - An obviously tailing peak is observed in a
chromatogram. The concentration of the standard
is decreased by a factor of 10 and the sample is
re-injected. The tailing looks about the same.
What can be concluded about the source of
tailing? List one other possible source of the
tailing. (added later)
19Chromatographic TheoryOptimization Some
Questions
- Why is it usually more difficult to improve the
separation factor (a) when there are a larger
number of analytes/contaminants? - Both using a longer column or using a column of
smaller H will improve resolutions. Which method
will lead to a better chromatogram? Why? - RS 0.93 and kB 2.7. What is the maximum RS
value just by changing kB? - An initial run of two standards at moderate
concentrations results in RS 1.9, kA 3.3 and
kB 4.0. Why might an analytical chemist and a
prep chemist change k in opposite directions?
20Gas ChromatographyOverview of Topics
- Comparison of mobile phases (Chapter 6)
- History, analyte stationary phase interaction
(Section 7.1) - Instrumentation (Section 7.2, 7.3)
- Stationary phase (Section 7.4)
- Temperature issues (Section 7.6)
21Gas ChromatographyComparison of Mobile Phases
- Two key differences between GC and LC
- No analyte mobile phase interaction in GC
- Temperature is routinely changed (and always
controlled) in GC - Effects of gases (vs. liquids)
- Much higher diffusivity (larger B term of van
Deemter equation but very small CM term) - Lower viscosity of gases (backpressure is not as
big an issue) - Much lower density (capacity of column is a big
issue with liquid samples) - Gases are compressible
22Gas ChromatographyCompressibility of Gases
- The volume flow rate will not be a constant along
a column because as the pressure drops, the
volume increases - There are various ways to calculate average flow
rates which we will not go into
23Gas ChromatographyAdvantages vs. HPLC
- Main practical advantage comes from high N values
(although H is usually larger) achieved with open
tubular columns. - Another advantage comes from being able to use
quite long columns (60 m vs. 250 mm for HPLC)
because backpressure is not a major issue - Other advantages have to do with instrument cost
and better detectors - Main disadvantage is for analysis of non-volatile
compounds
24Gas ChromatographyDevelopment and Theory
- Initially, GC was developed to improve upon
fractional distillations - In fractional distillations, the liquid at each
plate is a mixture of analytes - In gas chromatography analytes are present, but
stationary phase interactions are dominant and
analyte X and Y generally dont interact
X
Y
Liquid (or solid) stationary phase interacts with
x and y
Y
X
Liquid at each plate is mixture of distillates
(only X and Y)
25Gas ChromatographyDevelopment and Theory
- Types of Columns
- Packed Columns
- Older type of column
- Both solid and liquid stationary phase
- Best column for preparatory GC and for use with
thermal conductivity detectors - Sometimes used for very specific applications
(low production volume less of an issue) - Open Tubular Columns
- More modern columns
- Much better analytical performance (large N
values) - Most common in wall coated format (WCOT)
- Variety of diameters (0.25 to 0.53 mm most
common) allow high resolution vs. easier
injection - Stationary phases are mainly bonded of varying
amounts of polarity - Good reliability
- Disadvantages harder to make and less capacity
26Gas ChromatographyDevelopment and Theory
- Retention of Compounds
- KC value depends on
- Volatility
- Polarity of analyte vs. polarity of stationary
phase - Measure of volatility
- Best measure is vapor pressure at temperature
- Boiling point temperature is used more frequently
- Depends on molecules size and polarity
- Polarity in separations
- Compounds of similar polarity as stationary phase
will be more retained than similar compounds of
different polarity if their boiling points are
the same (ether vs. acetone example)
27Gas ChromatographyDevelopment and Theory
- Application of GC
- Gas samples
- Somewhat different equipment (injector and oven
range) is needed vs. liquid samples - Liquid samples
- Compounds must be volatile (plus small amounts of
non-volatile interferences) - Compounds must be stable at GC temperatures
- Separations are better for less polar compounds
- Issues occur for very volatile and low volatility
samples (due to min and max temperatures)
28Gas ChromatographyDevelopment and Theory
- Application of GC
- Extension to non-volatile, thermally labile
compounds - Derivatization example fatty acids are highly
polar and do not produce narrow peak with
non-polar columns, but they can be reacted to
produce fatty acid methyl ester (same reaction
used to produce biodiesel) that are volatile and
stable - Pyrolysis GC non-volatile samples are heated
and breakdown products are measured by GC. This
give information about compounds building
blocks
29Gas ChromatographyStationary Phase
- Selection of stationary phase affects k and a
values - Main concerns of stationary phase are polarity,
functional groups, maximum operating temperature,
and column bleed (loss of stationary phase due to
decomposition) - More polar columns suffer from lower maximum
temperatures and greater column bleed
Type Functional Groups Polarity
OV-1 methyl Non-polar
OV-17 50 methyl/50 phenyl Somewhat polar
OV-225 Cyanopropyl, methyl, and phenyl More polar
carbowax Ether groups polar
30GC InstrumentationMobile Phase
- Since there is no mobile phase analyte
interaction in GC, why does the mobile phase
matter? - Affects diffusion
- Smallest MW gases diffuse faster
- van Deemter B term at low flow rates (fast is
worse) and C term at higher flow rates (fast is
better) - Hmin not affected much, but umin affected by gas
chosen - Smallest MW allows fastest runs at min. H
- Detector requirements
- He is most common (inert, safe gas with high
diffusivity for better efficiency at high flow
rate) - H2 also can be used with even better efficiency,
but is less safe
CO2 min
H2 min
31GC InstrumentationSome Questions
- If a set of compounds in a sample could be
analyzed by GC or HPLC what would be two reasons
for picking GC? - What is a concern in analyzing a liquid sample
that has numerous highly volatile compounds? - In the case of the situation in question 2, would
you want a column with the stationary similar to
or different from the polarity of the analytes? - What is one way in which low volatility samples
can be analyzed by GC? - In response to high He prices, a lab director
says that no more He can be purchased. Would you
want to use Ne or N2? (assuming reasonable prices
for both of those gases)? What other change
would be needed to get reasonable separations
with Ne or N2 carrier gases? - How is the retention of polar compounds affected
by switching from He to H2 as a carrier gas?
32GC InstrumentationFlow Control
- Flow can be controlled by regulating inlet
pressure (either constant pressure or
compensation for constant linear velocity). - Equipment consists of valves for regulating
pressure (constant pressure) in older instruments
or electronic pressure control (solenoid valve
opens or closes in response to pressure). - Flow rate is typically checked at detector using
bubble meter.
Pressure Transducer
Solenoid valve
Soap film
soap
33GC InstrumentationSample Injection
- Several types of injectors are available and
choice of injector depends on sample phase,
analyte concentration, and other sample
properties - The most common injectors are designed for
liquids (but can be used for gases) - Injectors for gases only can be used for gases
- Liquids require much smaller volumes (1 µL, a
typical liquid injection volume, is equivalent to
1 mL after evaporation) and column overloading
is common - Column overloading is most common with narrow
diameter OT columns and least common with packed
columns - Most injectors are heated (except on-column)
34GC InstrumentationSample Injection Gas Samples
6 port valve
- Fixed Loop Injectors
- A loop of fixed volume is filled with a gas
- The injection valve is twisted so that the mobile
phase pushes the gases in the loop into the
column - Very similar to most common injections in HPLC
(Covered later) - Very reproducible injection
He in
To GC column
Waste
Gas sample in
INJECT POSITION
LOAD POSITION
35GC InstrumentationSample Injection Gas Samples
- Specialized Injectors (Fixed loop injectors with
trapping capability) - Best for trace analysis
- In place of loop is a trap (adsorbant or cold
trap) so that all gas sent into loop gets
trapped, then injected - These allow injection of greater volumes but may
require removal of interferents (oxygen, water)
and require better quantitative control of gases
(careful volume or pressure monitoring) - Thermal trapping (cool to trap, then hot to
desorb) can increase efficiency - Other ways to inject gas samples (using injectors
designed for liquids) - Direct syringe injection (samples at higher
concentrations) - Solid phase microextraction (SPME with fibers
exposed to gas samples)