Title: Aucun titre de diapositive
1The Calorimeter Upgrade
- Run I ? Run II
- increase of the Beam Energy
- E (TeV) 1.8 ? 2.0
- increase of the Luminosity
- L (cm-2 sec-1) 2x1031 ? 2x1032
- modification of bunch structure
- decrease of interaction time
- Dt 3.6 ms ? 396 (132) ns
this gap is too small to form trigger and
sample baseline
Run II No modification of the calorimeter
structure New Electronics due to timing
constraints
- DØ Calorimeter
- Uranium / liquid Argon
- 55000 channels
- ? lt4
- D? ? D? 0.1 ? 0.1
- ?(em) 15 / ?E
- ?(had) 50 / ?E
- Excellent Run I performances
- low noise 10 MeV
- linearity lt 0.5
SCA analog delay gt 4 ms, alternate
new low noise preamp driver
- Signal shaping with peak
- sampling time at 400 ns
- Analog buffering for 4ms
- before L1 trigger decision
Trig. sum
Bank 0
Electronic Calibration
SCA (48 deep)
SCA (48 deep)
x 1
Filter/ Shaper
Output Buffer
Preamp/ Driver
SCA
BLS
x 8
SCA (48 deep)
? Replacement of Preamps, Shapers, BLS ?
addition of SCA ? new calibration.
SCA (48 deep)
Bank 1
Additional buffering for L2 L3
Replace cables for impedance match
Shorter shaping 400ns
Switched Capacitor Array
Preamplifiers
- Not designed for simultaneous
- read and write operations
- Calorimeter preamp
- hybrid on ceramic
- 48 preamps on
- one motherboard
- similar to previous version except
- Dual FET frontend
- Compensation for detector capacitance
- Faster recovery time
? two SCA banks alternate reading and writing
- approximately deadtimeless
- to L1 rate up to 100kHz
- Only 12 bit dynamic range
driver
preamp
New output driver for terminated signal
transmission
FET
? low and high gain path for each readout
channel (X8/X1) ? maintain 15 bit dynamic range
Electronic Calibration System
- Expected Run II Noise contributions
- (compared to Run I)
- Electronics noise ? 1.6
- increases due to shorter shaping times
- (2 ms to 400 ns as ?t)
- decrease due to lower noise preamp
- (2 FET input as 1/ ? 2)
- Uranium noise ? 2.3
- decrease due to shorter shaping times (as ? t)
- Pile-up noise ? 1.3
- increase due to luminosity (as ? L)
- decrease due to shorter shaping times (as ? t)
- fast pulse similar to physics signal
- optimized calibration for various
- detector capacitances possible
- precision better than 1
Pulser linearity
Frédéric.Machefert_at_in2p3.fr
Pulse height
Run II calorimeter electronics performances
equivalent to Run I for a 10 times higher
luminosity .