Title: Folie 1
1 FLC Group Test-beam Studies of the
Laser-Wire Detector 13 September
2006 Maximilian Micheler Supervisor Freddy
Poirier
2Introduction
The goal of the Test-Beam Studies is to measure
the performance of the lead-tungstate detector
used at the Laser-Wire experiment at PETRA
II ? Calibration, for the use of the detector
at the Laser-Wire experiment ? Resolution,
mainly for simulation purposes
3Experimental setup
- Detector is placed in line with the electron beam
as to achieve an EM shower in the centre of the
lead-tungstate crystal. (achieved by mounting the
detector on an adjustable table) - Hamamatsu R6236 photomultiplier tube (54mmx54mm
active photo cathode surface) for detection of
the energy dissipation in the crystal
- Two scintillators (one horizontal and one
vertical scintillator) were placed in front of
the detector and connected to a coincidence unit
to only record signals for electrons which
entered the crystal in the centre - The electron beam energies were selected by
changing the currents in the magnets using a
computer with a provided software in the
Test-beam 24 control room.
4Electronics
- Scintillator signal processed using
- Amplifier
- Discriminator
- Coincidence unit
- TTL signal converter
- TTL signal used for triggering integration of PMT
output signal.
- Using a Computer (ADC Break-out box in the
figure) and an integrator card to integrate the
PMT signal and to record the area. - The setup of the test-beam studies was chosen to
have the same conditions as in the Laser-Wire
Hut - Identical electronics for the read-out system,
i.e. ADC box and integrator card - Similar timing for read-out system
5Data Acquisition
- Signal from PMT is a negative voltage peak.
- Integrator card integrates this peak and outputs
the area as a positive voltage which is recorded
by the computer. - For every single beam energy and PMT supply
voltage around 1000 integrator card voltages have
been measured.
- The 1000 area measurements were plotted on a
histogram and fitted with a Gaussian function - The two graphs are taken for a PMT supply voltage
of 1115V at a beam energy of 3.6GeV - Mean, Standard deviation, and Errors on the
Standard deviation from Gaussian function
6Detector Performance
- Calibration plot the mean of the integrated
signal against the corresponding beam energy for
the 4 different PMT supply voltages - This calibration plot was expected to show a
linear dependence of the integrated signal on the
beam energy - Linear fittings of the individual data sets show
a non-linear behaviour
Calibration plot
- Resolution plot the PMT normalised resolution
(Gaussian width/integrated signal) against the
beam energy - Resolution is expected to decrease with
increasing beam energy. - At a PMT supply voltage of 1300V this is not the
case
Resolution plot
7Control Tests
- Possible reasons why the results differ from
expectation - Saturation of PMT
- Cutting the signal
- ? Incorrect delay
- ?Incorrect integration width
- Saturation of the read-out system
- Investigation of the raw PMT output signal
recorded from the oscilloscope (as shown). The
following quantities were directly calculated
from the raw signal - Signal amplitude (difference between constant
base line and minimum voltage) - Signal area (sum of the relative signal w.r.t.
the base line within the integration range)
8Control Tests PMT saturation effects
- Plot of signal amplitude against signal area
shows a linear dependence - Fit y 0.0274x 0.0474
- where y represents the signal amplitude and x the
signal area. - Plot of signal amplitude against beam energy
shows a linear dependence - Fit y 0.404x 0.076
- where y represents the signal amplitude and x the
beam energy. - Therefore the signal area also shows a linear
dependence on the beam energy - ? No detector saturation while increasing the
beam energy
9Control Tests Cut-off effects
- 1µs integration delay w.r.t. the TTL trigger,
2.4µs integration width - The two graphs show a integration with an
asymmetric cut-off of the signal from the tail
and the front of the PMT raw signal,
respectively. - Signal area decreases due to cut-off. However,
the plots of the signal area against the beam
energy for the different cut-off points are still
linear. - ? Cut-off effects do not seem to be responsible
for the non-linear characteristics of the
integrated signal - ?No cut-off effects due to integration width as
width is sufficiently long enough to detect
entire signal (width approx. 110ns)
tail
front
10Control Tests Read-out system saturation
At the present time, the read-out system is under
study to check how it performs with voltages
higher than it is designed for.
Conclusions
- Resolution plot and Calibration plot achieved for
the Laser-Wire detector - Results used for calibration of the signal at the
Laser-Wire experiment - After the calibration the data from the
Laser-Wire experiment will be compared with
previous simulations